  • Name: Portable Oxygen Cylinder Filling System
  • Date: 2015-08-06
  • Views : 96

Friends Recommend

The oxygen filling system consists of a high pressure oil-free compressor and a PSA oxygen concentrator. Through the compressor, the oxygen generated by the concentrator will be filled into different size of oxygen cylinders. By using this oxygen filling equipment, you can generate and store oxygen whenever and wherever you like.
Application
This oxygen filling system can used for medical, homecare, industrial and other applications where delivery of oxygen cylinders is not convenient. It is especially suitable for EMS oxygen supply for life rescue in disaster areas as well as health care in underground or plateau areas with low oxygen concentration.
Features
1. Absolutely dry, clean and oil-free.
2. High compression ratio. The service pressure is up to 138bar (2000psi/13.8Mpa)
3. Efficient filling. It takes about 6 hours to fill a M265 10-liter oxygen cylinder using our oxygen filling system.
4. Small size and light weight. Easy and safe to operate.
5. Simple, reliable and easy maintenance.
6. High flexibility. This oxygen filling equiment can be used separately or expanded into a larger system through parallel connection to meet the demand for large amount of oxygen consumption.
7. Energy-saving. Our oxygen filling device has low power consumption.
8. Cost-effective.

 

Parameters

 

Model

JAY-10

JAY-20

Rated power (VA)

550W

1100W

Operation voltage (V/Hz)

AC230V (± 10%); 50Hz (±1Hz)

Oxygen flow (L/min)

0-10

0- 20

Oxygen concentration

93% (± 3% )

Outlet pressure (Mpa)

0.04-0.07

Alarm

Power failure; low/high pressure; temperature (optional); low concentration (optional); maintenance reminding after 3000 hours' operation (optional)

Noise level (dB)

≤ 55

≤ 55

LCD display

Accumulating timing; present timing; timing;

Large LCD display (optional)

Pressure indication (accuracy: 0.001MPa);
Accumulated working time (range: 0-10000hours);
Present working time (accuracy:1 minute);
Preset time (accuracy: 1 minute)
Optional: temperature indication (accuracy: 0.1);
Oxygen concentration indicationMaintenance reminding

Electrical safety classification

Class II Type B

Product category

Class II a

Net weight (kg)

28 (plastic shell)/30 (iron shell)

60

Dimension (mm)

365×375×600 ( plastic shell)380×380×560 (iron shell)

700×390×530

Low purity alarm (Optional)

When oxygen concentration is 85% or above, the green lamp is on; when oxygen concentration isbelow 85%, the red lamp is on, indicating low concentration

Fuse

F7A/250V

 

High Pressure Oil-Free Compressor Specification

 

 

 

10L

20L

Electrical requirement

C150-500

C150-800

Power consumption

500W on average

1000W on average

Inlet pressure

6-20psi (0.04-0.01Mpa)

6-20 psi (0.04-0.01Mpa)

Max. discharge pressure

2000psi (13.8Mpa)

2000psi (13.8Mpa)

Flow rate

5-10LPM

15-20LPM

Machine dimension (W×D×H)

38cm×62cm×63cm

38cm×62cm×63cm

Net weight

52.5kg

64kg

 

 

 

 

 

 

 

 

New High Pressure Oil-Free Compressor Specification

 

 

Operating temperature

5-35

Operating humidity range

15-95%RH (No condensation)

Operating altitude

0-2000m

Storage temperature range

-40 -+70

Storage humidity range

10%-95%RH (No condensation)

Working voltage

115V±10% /60Hz (230V±10%/50Hz)

Power

700W on average

Classification

Category Model B

Inlet airflow

5-10LPM

Inlet pressure

7-15psi

Filling pressure

0psi-2000 (±200psi)

Dimension (L×W×H)

510mm×510mm×800mm

Weight

About 60kg

 

Filling Time for Various Types of Cylinders with Oxygen Filling System

 

 

Cylinder model

Cylinder capacity

Oxygen storage

Air flow

Input air pressure

Output air pressure

Cylinder filling time

 

L

L

LPM

psi

Mpa

psi

Mpa

 

M004

0.7

98

10/20

15/20

0.1/0.14

2000

13.8

12L: 5mins

M006

1.0

140

10/20

15/20

0.1/0.14

2000

13.8

16L: 7mins

M009

1.7

238

10/20

15/20

0.1/0.14

2000

13.8

25L: 12mins

M265

40

6800

10/20

15/20

0.1/0.14

2000

13.8

<12L: 6hours

 

Key words :

download attachments :

  To be closed automatically in X seconds

View my favorites

No related products Information

 
Content:
* Entered characters:0 Word
Less than or equal500Character
Verification code:
   

There is no corresponding message.